Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MNA-420 ENC HDMI Recalled by Barco N.V. Due to We have identified a potential issue with the...

Date: April 24, 2024
Company: Barco N.V.
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Barco N.V. directly.

Affected Products

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Quantity: 88 units

Why Was This Recalled?

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Where Was This Sold?

This product was distributed to 1 state: FL

Affected (1 state)Not affected

About Barco N.V.

Barco N.V. has 5 total recalls tracked by RecallDetector.

Related Recalls