Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product Recalled by Autochair Due to Recalling firm was informed that as a result...

Date: March 7, 2018
Company: Autochair
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Autochair directly.

Affected Products

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Why Was This Recalled?

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Where Was This Sold?

This product was distributed to 1 state: PA