Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MONARCH Platform Recalled by Auris Health, Inc Due to Device for bronchoscopic visualization, patient airway access has...

Date: January 21, 2026
Company: Auris Health, Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Auris Health, Inc directly.

Affected Products

MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software

Quantity: 173

Why Was This Recalled?

Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Auris Health, Inc

Auris Health, Inc has 2 total recalls tracked by RecallDetector.

Related Recalls