Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

AMSORB PLUS PREFILLED G-CAN 1.0L Recalled by Armstrong Medical Services Limited Due to Some devices may be associated with high and...

Date: August 5, 2021
Company: Armstrong Medical Services Limited
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Armstrong Medical Services Limited directly.

Affected Products

AMSORB PLUS PREFILLED G-CAN 1.0L; Carbon dioxide absorbent canister; Product codes AMAB3801 and AMAB3801GE. Used in anesthesia.

Quantity: 214,032 boxes

Why Was This Recalled?

Some devices may be associated with high and unexpected resistance to gas flow during clinical use, impacting or preventing adequate ventilation of an anaesthetised patient, which could cause a delay in treatment. Not all defective canisters will be identified by the pre-use test.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Armstrong Medical Services Limited

Armstrong Medical Services Limited has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report