Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

NOMAD Pro Packaged X-Ray System Recalled by Aribex Inc Due to Firm received customer complaints where end users reported...

Date: July 18, 2016
Company: Aribex Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Aribex Inc directly.

Affected Products

NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD Pro Packaged X-Ray System - Europe, Part Number 08500013; NOMAD Pro X-Ray System, Remote Config., Part Number 08500017; NOMAD Pro Packaged X-Ray System JP, Part Number 08500019

Quantity: 16,871

Why Was This Recalled?

Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Aribex Inc

Aribex Inc has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report