Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brand Name Burn Navigator¿ Model Numbers: Catalog 1120 Recalled by Arcos, Inc Due to Battery packs may heat up abnormally and ignite...

Date: March 16, 2015
Company: Arcos, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Arcos, Inc directly.

Affected Products

Brand Name Burn Navigator¿ Model Numbers: Catalog 1120, Burn Navigator¿ H2 Catalog 1125, Burn Navigator¿ H2 Transport Package Burn Navigator User Manual Manual-1120 Burn Navigator User's Manual, Rev. F Battery: Panasonic Model CF-H2 Part Number CMP-1523

Quantity: 28

Why Was This Recalled?

Battery packs may heat up abnormally and ignite in rare instance.

Where Was This Sold?

This product was distributed to 6 states: AZ, LA, MD, OR, TX, WA

Affected (6 states)Not affected

About Arcos, Inc

Arcos, Inc has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report