Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
NAMIC CONVENIENCE KIT RENOWN REGIONAL MEDICAL CENTER Recalled by AngioDynamics Inc. Due to Procedure Products initiated the recall due to the...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact AngioDynamics Inc. directly.
Affected Products
NAMIC CONVENIENCE KIT RENOWN REGIONAL MEDICAL CENTER, REF Catalog No. 65230029, UPN Product No. H965652300290, Product is packed as a box of 30 pieces, box label contains UPN Product No. H965652300291 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire
Quantity: 100 kits
Why Was This Recalled?
Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About AngioDynamics Inc.
AngioDynamics Inc. has 4 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report