Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MINI STICK MAX 4F X 10 CM STIFF .018 SS/PD Recalled by Angiodynamics, Inc. Due to Non-conformance may prevent the guidewire from passing through...

Date: January 30, 2024
Company: Angiodynamics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Angiodynamics, Inc. directly.

Affected Products

MINI STICK MAX 4F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-752

Quantity: 58 units

Why Was This Recalled?

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Angiodynamics, Inc.

Angiodynamics, Inc. has 74 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report