Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Custom procedural convenience kits and trays Recalled by American Contract Systems, Inc. Due to During an internal investigation, ACS identified that several...

Date: November 29, 2023
Company: American Contract Systems, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact American Contract Systems, Inc. directly.

Affected Products

Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, kit number HNLP12J; j) MIDLINE CHANGE KIT, kit number HSMD20E; k) Minor General, kit number JSMI10I; l) ENDO VASCULAR PACK, kit number LVEV61; m) MAJOR PACK, kit number LVMJ21; n) HEAD AND NECK PACK, kit number MMHN37F; o) KIT, MINOR U BAR, kit number MMNU34E; p) KIT, MAJOR UNIVERSAL, kit number MMUN28E; q) VAG PLASTY PACK, kit number OSVP16; r) DR L MAJOR, kit number SALM40H; s) DR L MAJOR, kit number SALM40I; t) MINOR BASIN SAH, kit number SAMB25E; u) MAJOR SAH, kit number SAMJ36G; v) MINOR PACK, kit number SMMN65; w) General Pack, kit number THGN59D; x) MINOR PACK, kit number TPMN15; y) MAJOR UNIVERSAL PACK, kit number TPUN48

Quantity: 7881 units

Why Was This Recalled?

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About American Contract Systems, Inc.

American Contract Systems, Inc. has 90 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report