Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Basic Biopsy Tray Recalled by American Contract Systems Inc. Due to Identified products were re-gassed (sterilized more than once)...

Date: September 4, 2025
Company: American Contract Systems Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact American Contract Systems Inc. directly.

Affected Products

Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PACK - 205956, ANLP25AJ LAPROSCOPY PACK - 205958, FYLA38E LAPCHOLAPPY PACK, IHDA88AH DIVINCI GYN URO, IHLC11AE LAP GENERAL PACK, MHLP45AJ LAPAROSCOPY PACK - 206005, MHRB88AO ROBOTIC PACK - 242561, RCLC51C LAP CHOLE RCH 0246969, SMLC53C LAP CHOLE PACK, UDLC17AE DSC LAP CHOLE PACK - 206062, UIRU84AE ROBOTIC UROLOGY PACK (PS 060571)

Quantity: 32,433 total

Why Was This Recalled?

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About American Contract Systems Inc.

American Contract Systems Inc. has 18 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report