Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
AMERICAN CONTRACT SYSTEMS Recalled by American Contract Systems Inc Due to During an internal investigation, ACS determined that the...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact American Contract Systems Inc directly.
Affected Products
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE PACK, Model/Item Number JKLC55H; 4) LAPAROSCOPY PACK, Model/Item Number LLLA21K; 5) GENERAL LAPAROSCOPY MPH, Model/Item Number MPGL15G;
Quantity: 1233 units
Why Was This Recalled?
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Where Was This Sold?
This product was distributed to 2 states: IL, LA
About American Contract Systems Inc
American Contract Systems Inc has 81 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report