Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
AMERICAN CONTRACT SYSTEMS Recalled by American Contract Systems Inc Due to During an internal investigation, ACS determined that the...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact American Contract Systems Inc directly.
Affected Products
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM NUMBER ANVA82AB; 3) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48O; 4) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48P; 5) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18F; 6) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18G; 7) CABG ACCESSORY PACK, MODEL/ITEM NUMBER HGCB98F; 8) CV SET UP PACK, MODEL/ITEM NUMBER HGCV05K; 9) OPEN HEART PACK B - COMPONENTS, MODEL/ITEM NUMBER HGOH99H; 10) STRADLE TABLE PACK - 301659, MODEL/ITEM NUMBER MHCV10AM; 11) MINOR VASCULAR PACK, MODEL/ITEM NUMBER MTVS73D; 12) AV GRAFT RCH 0246952, MODEL/ITEM NUMBER RCAV40C; 13) OPEN HEART PART 1 RCH 0247159, MODEL/ITEM NUMBER RCOH61G; 14) CARDIO 612968, MODEL/ITEM NUMBER TNCA08T; 15) CV B PACK, MODEL/ITEM NUMBER TNCB10AM; 16) A-V FISTULA - 206042, MODEL/ITEM NUMBER UDAV16AD; 17) CAROTID PACK - 206045, MODEL/ITEM NUMBER UDCP24AF; 18) FEM POP PACK - 206052, MODEL/ITEM NUMBER UDFP36AH; 19) CARDIOVASCULAR SUPPLY (PS 907285), MODEL/ITEM NUMBER UICD62AN; 20) SFCH PEDS CARDIAC SUPPLY PACK (PS131653), MODEL/ITEM NUMBER UICS88S; 21) SFCH ECMO PACK (PS 107116), MODEL/ITEM NUMBER UIPD03Q; 22) THORACIC SETUP (PS 035700), MODEL/ITEM NUMBER UITC53AL; 23) VASCULAR PACK, MODEL/ITEM NUMBER UIVA75J; 24) CABG ACCESSORY PACK 175619, MODEL/ITEM NUMBER UMCA14E; 25) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42E; 26) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42F; 27) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42G.
Quantity: 31506 units
Why Was This Recalled?
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Where Was This Sold?
This product was distributed to 2 states: IL, LA
About American Contract Systems Inc
American Contract Systems Inc has 81 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report