Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ambu VivaSight 2 DLT Recalled by Ambu Inc. Due to Firm has received complaints on the device concerning...

Date: June 3, 2022
Company: Ambu Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ambu Inc. directly.

Affected Products

Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000, 412371000, 412391000, 412411000

Quantity: 12,921

Why Was This Recalled?

Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ambu Inc.

Ambu Inc. has 7 total recalls tracked by RecallDetector.

Related Recalls