Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

A&E Medical Thorecon Plating System Kits Model No. 94-1700-10. For Recalled by Alto Development Corp Due to Some Thorecon disposable tensioners supplied in different Plating...

Date: September 10, 2019
Company: Alto Development Corp
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Alto Development Corp directly.

Affected Products

A&E Medical Thorecon Plating System Kits Model No. 94-1700-10. For fracture fixation or reconstruction of the sternum.

Quantity: 209 total

Why Was This Recalled?

Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which occludes the through-channel for cable tensioning, rendering the tensioner inoperable.

Where Was This Sold?

This product was distributed to 16 states: AL, CA, FL, GA, IL, LA, MS, MO, NM, OK, SC, TX, UT, VT, WA, WI