Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Triage Drugs of Abuse Plus TCA 25 Test Kit Recalled by Alere San Diego, Inc. Due to The recalled lots have demonstrated unexpected false positive...

Date: November 14, 2017
Company: Alere San Diego, Inc.
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Alere San Diego, Inc. directly.

Affected Products

Triage Drugs of Abuse Plus TCA 25 Test Kit, Model Numbers 92000, 92590, 92591; 25 individually pouched devices, 1 bottle of Wash Solution, 1 Pipette and 50 Pipette Tips in a labeled kit box. Product Usage: The Alere Triage Drugs of Abuse Panel plus Tricyclic Antidepressants uses eight discrete competitive immunoassay procedures for the simultaneous detection and identification of the major drugs and/or the urinary metabolites of eight different drug classes. The use of monoclonal antibodies that are specific for the drugs and/or metabolites of the eight drug classes ensures a high degree of sensitivity and specificity.

Quantity: 2940 units

Why Was This Recalled?

The recalled lots have demonstrated unexpected false positive THC results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Alere San Diego, Inc.

Alere San Diego, Inc. has 49 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report