Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Alere INRatio2 PT/INR Professional Monitoring System Model Number: 55128A Product Recalled by Alere San Diego, Inc. Due to Alere San Diego is initiating a voluntary recall...

Date: July 11, 2016
Company: Alere San Diego, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Alere San Diego, Inc. directly.

Affected Products

Alere INRatio2 PT/INR Professional Monitoring System Model Number: 55128A Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes.

Why Was This Recalled?

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Where Was This Sold?

This product was distributed nationwide across all 50 states.