Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.
Alere INRatio PT/INR Test Strips Recalled by Alere San Diego, Inc. Due to In certain cases, an INRatio PT/INR Monitor system...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Alere San Diego, Inc. directly.
Affected Products
Alere INRatio PT/INR Test Strips, Alere INRatio PT/INR System Professional - 0100071 Alere INRatio PT/INR Test Strips, Box of 12 - 0100139 Alere INRatio PT/INR Test Strips, Box of 48 The test strips are packaged in individually pouched test strips in a labeled box with 12 or 48 test strips per box. In vitro diagnostic system that provides a quantitative prothrombin time result, expressed in seconds and as an International normalized ratio (INR).
Quantity: 77,342 units (12 test strips/box) and 308,947 units (48 test strips/box)
Why Was This Recalled?
In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Alere San Diego, Inc.
Alere San Diego, Inc. has 49 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report