Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Phaco Standalone Tips Recalled by Alcon Research, LTD. Due to Reports of patient injury during use in procedures...

Date: June 28, 2023
Company: Alcon Research, LTD.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Alcon Research, LTD. directly.

Affected Products

Phaco Standalone Tips

Quantity: 444,372 units

Why Was This Recalled?

Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Alcon Research, LTD.

Alcon Research, LTD. has 14 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report