Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Constellation Vision System Recalled by Alcon Research, LLC Due to Ophthalmic microsurgical vision systems have a new Printed...

Date: August 18, 2023
Company: Alcon Research, LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Alcon Research, LLC directly.

Affected Products

Constellation Vision System, REF: 8065751150

Quantity: 3

Why Was This Recalled?

Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.

Where Was This Sold?

This product was distributed to 2 states: GA, IL

Affected (2 states)Not affected

About Alcon Research, LLC

Alcon Research, LLC has 9 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report