Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Airways Development LLC WaterPAP Positive Airway Pressure Device Reference #WP-7700. Recalled by Airways Development LLC Due to Airways Development LLC has received a complaint about...

Date: September 24, 2015
Company: Airways Development LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Airways Development LLC directly.

Affected Products

Airways Development LLC WaterPAP Positive Airway Pressure Device Reference #WP-7700.

Quantity: 1,464 units

Why Was This Recalled?

Airways Development LLC has received a complaint about a canister leaking from WaterPAP Lot 8605A. During a review with their supplier it could not be determined if this was an isolated incident or if the whole lot is impacted.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Airways Development LLC

Airways Development LLC has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report