Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ENDURO SPECIAL KEY TIB.LOCK.RING F1/10MM product code NP462R used with Recalled by Aesculap Implant Systems LLC Due to Nonfunctional key due to the potential for damage...

Date: November 12, 2019
Company: Aesculap Implant Systems LLC
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Aesculap Implant Systems LLC directly.

Affected Products

ENDURO SPECIAL KEY TIB.LOCK.RING F1/10MM product code NP462R used with EnduRo Total Knee System. instrument used during orthopedic implant surgery.

Quantity: 67

Why Was This Recalled?

Nonfunctional key due to the potential for damage to the key pins cause by force of use during assembly of the tibia locking ring.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Aesculap Implant Systems LLC

Aesculap Implant Systems LLC has 61 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report