Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

EVOTECH Endoscope Cleaner and Reprocessing System 208V Recalled by Advanced Sterilization Products Due to Advanced Sterilization Products (ASP) is recalling the EVOTECH...

Date: May 11, 2015
Company: Advanced Sterilization Products
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Advanced Sterilization Products directly.

Affected Products

EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.

Quantity: 825 units

Why Was This Recalled?

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Advanced Sterilization Products

Advanced Sterilization Products has 36 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report