Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Disposable Drill Kit which includes two drill bits Recalled by Ad-Tech Medical Instrument Corporation Due to There is a possibility that DDK2-2.4-30X Disposable Drill...

Date: December 20, 2017
Company: Ad-Tech Medical Instrument Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ad-Tech Medical Instrument Corporation directly.

Affected Products

Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.

Quantity: 30 devices

Why Was This Recalled?

There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ad-Tech Medical Instrument Corporation

Ad-Tech Medical Instrument Corporation has 17 total recalls tracked by RecallDetector.

Related Recalls