Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AD-TECH Spencer Probe Depth Electrode Recalled by Ad-Tech Medical Instrument Corporation Due to Ad-Tech is recalling these A-Style SD Depth Electrodes...

Date: September 19, 2024
Company: Ad-Tech Medical Instrument Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ad-Tech Medical Instrument Corporation directly.

Affected Products

AD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth Electrode

Quantity: 4 units

Why Was This Recalled?

Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ad-Tech Medical Instrument Corporation

Ad-Tech Medical Instrument Corporation has 17 total recalls tracked by RecallDetector.

Related Recalls