Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw Recalled by Ad-Tech Medical Instrument Corporation Due to The diameter of the drill sleeve guide raw...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Ad-Tech Medical Instrument Corporation directly.
Affected Products
AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from vendor 00-0001062., AD0021502, AD0021628. Product Usage: The drill sleeve guides are accessories to the Ad-Tech Cranial Drill Bits. The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement. The DSG-6.3-090-2.4N drill sleeve guides are intended to be used with the 2.4mm Cranial Drill Bit diameter.
Quantity: 65 Total
Why Was This Recalled?
The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Ad-Tech Medical Instrument Corporation
Ad-Tech Medical Instrument Corporation has 17 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report