Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ashland(TM) Box Film Ball Cube II EBT3 Recalled by Accuray Incorporated Due to the Ball Cube II phantom calibration film was...

Date: August 31, 2018
Company: Accuray Incorporated
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Accuray Incorporated directly.

Affected Products

Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems

Quantity: 259 units

Why Was This Recalled?

the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Accuray Incorporated

Accuray Incorporated has 20 total recalls tracked by RecallDetector.