Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
VerifyNow System Recalled by Accumetrics Inc Due to The recall was initiated by Accumetrics because of...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Accumetrics Inc directly.
Affected Products
VerifyNow System, part #85005-6H Product Usage: The VerifyNow P2Y12 assay device is a whole blood assay used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. The VerifyNow Systems intended for use with human whole blood and verify now test devices. The system should be operated by healthcare professionals trained on use of the system and in accordance with institution and policies and procedures. Accumetrics respresentatives will assist your institution in the installation of the system and operations of the operator.
Quantity: 1,167 units total
Why Was This Recalled?
The recall was initiated by Accumetrics because of a field correction to the VerifyNow System; specifically the results reported when running the VerifyNow P2Y12 Test. Currently the VerifyNow P2Y12 Test reports three values: PRU, BASE and % Inhibition. The change implemented is the elimination of the BASE and % Inhibition results. There will be no change to the reporting of the PRU result whic
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Accumetrics Inc
Accumetrics Inc has 4 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report