Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
VerifyNow IIb/IIIa Test Recalled by Accumetrics Inc Due to Accumetrics is recalling the VerifyNow IIb/IIIa 10-Test because...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Accumetrics Inc directly.
Affected Products
VerifyNow IIb/IIIa Test, 10-Test Kit, Catalog No. 85310, 10 tests/box; 690 total tests. The VerifyNow IIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow lIb/IlIa Test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician. The VerifyNow System is a turbidimetric based optical detection system which measures platelet-induced aggregation. The system consists of an instrument, a disposable test device and quality control materials. The VerifyNow lIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide.
Quantity: 69 units
Why Was This Recalled?
Accumetrics is recalling the VerifyNow IIb/IIIa 10-Test because it contains incorrect information for the amount of time in which to perform the test following collection of the patient sample. The correct instruction is to perform the test within 15 minutes of collecting the patient sample.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Accumetrics Inc
Accumetrics Inc has 4 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report