Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Abbott Point of Care i-Stat cTnI cartridge. The i-Stat cardiac Recalled by Abbott Point of Care Canada Limited Due to APOC has determined that 10 boxes of i-Stat...

Date: November 1, 2012
Company: Abbott Point of Care Canada Limited
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Point of Care Canada Limited directly.

Affected Products

Abbott Point of Care i-Stat cTnI cartridge. The i-Stat cardiac troponin I (cTnI) test is an in vitro diagnostic test for the quantitive measurement of cardiac troponin I (cTnI) in whole blood or plasma. Measurements of cardiac troponin I are used in the diagnosis and treatment of myocardial infarction and as an aid in the risk stratification of patients with acute coronary syndromes with respect to the relative risk of mortality.

Quantity: 560 boxes (25 cartridges in each box)

Why Was This Recalled?

APOC has determined that 10 boxes of i-Stat cTnI cartridges from lot P12179 have an incorrect barcode applied to the cartridge portion pack.

Where Was This Sold?

Worldwide Distribution.