Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Alinity m System Recalled by Abbott Molecular, Inc. Due to Abbott has identified potential performance issues for the...

Date: October 12, 2022
Company: Abbott Molecular, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Molecular, Inc. directly.

Affected Products

Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use

Quantity: 883 units

Why Was This Recalled?

Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and locked. It was found that the software will cause the flapper door to stay in the closed position when the drawer is closed and locked if the flapper had previously been moved out of position. 2) Sample preparation drawer #1 barcode information is used instead of sample prep drawer #2 when the scanned data is not sent to the System Control Center before the next bottle barcode in sample prep drawer#2 is scanned. 3) Under a specific condition, while the system is processing tests and a new test request is made, when the level of the bulk solution is too low to process a test, the software would error stop the system and try to complete all in-process tests. An internal counter is reset which can cause reaction vessels (RVs) already present in an Amplification Detection Unit (or amp detect) to remain there and not be moved to waste so that more RVs can be placed into the amp detect for processing. 4) 4 to 6 replicates of the same auto calibrator orders can be run with 2 different sets of calibrator materials on the same rack. It was discovered during internal testing, when running 4 to 6 replicates of a calibrator, it is possible for the user to use 2 separate lots of material which is not detected by the System Control Center (SCC). If this were to occur, the calibration curve would be created using 2 separate material lots. In normal use scenario, this curve would be made using only 1 lot of material.

Where Was This Sold?

Worldwide distribution.