Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Abbott Alinity m STI AMP Kit Recalled by Abbott Molecular, Inc. Due to Abbott has identified an increase of incidences regarding...

Date: September 12, 2024
Company: Abbott Molecular, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Molecular, Inc. directly.

Affected Products

Abbott Alinity m STI AMP Kit, used with the Alinity m System, product codes: a) REF 09N17-095; b) REF 09N17-090; c) REF 09N17-091;

Quantity: 15068 units

Why Was This Recalled?

Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Abbott Molecular, Inc.

Abbott Molecular, Inc. has 30 total recalls tracked by RecallDetector.

Related Recalls