Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Radiofrequency Grounding Pad Recalled by Abbott Medical Due to Affected lots were manufactured with the protective release...

Date: October 14, 2019
Company: Abbott Medical
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Medical directly.

Affected Products

Radiofrequency Grounding Pad, REF RF-DGP-L

Quantity: 8940 devices

Why Was This Recalled?

Affected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere with patient skin contact and lead to uneven heating.

Where Was This Sold?

This product was distributed nationwide across all 50 states.