Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

CLASSIC Radiofrequency Cannula - Curved Recalled by Abbott Medical Due to A manufacturing error caused a single lot of...

Date: November 11, 2019
Company: Abbott Medical
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Medical directly.

Affected Products

CLASSIC Radiofrequency Cannula - Curved, 10 cm, SMK Compatible Electrode, REF C-1005-S - Product Usage: Disposable Radiofrequency Cannulae are provided in either straight, curved, or curved blunt geometries with a stylet inserted in the cannula. Cannulae are insulated except for the active tip. The active tip determines the size of the radiofrequency lesion.

Quantity: 470 devices

Why Was This Recalled?

A manufacturing error caused a single lot of 10mm active tip length Cannulas to be labeled as 5mm active tip length cannulas.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Abbott Medical

Abbott Medical has 42 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report