Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

CardioMEMS HF System Hospital and Patient Electronics Units: (a) Patient Recalled by Abbott Laboratories, Inc. Due to Abbott is advising customers that a small number...

Date: June 14, 2018
Company: Abbott Laboratories, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Laboratories, Inc. directly.

Affected Products

CardioMEMS HF System Hospital and Patient Electronics Units: (a) Patient Electronics System, Model CM1100 (b) Hospital Electronics System, Model CM3000

Quantity: 2521 units

Why Was This Recalled?

Abbott is advising customers that a small number of CardioMEMS(R) Hospital Electronics Systems (Model CM3000) and Patient Electronics Systems (Model CM1100) may deliver a system error, known as Error 5. While this error message is intended to present if the electronics system exceeds a certain temperature, these units may deliver a false Error 5 message due to an incorrectly configured component within the device electronics.

Where Was This Sold?

US, Canada, Netherlands, Germany, Great Britain, France, Ireland, Belgium