Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARCHITECT c4000 part number 02P24 Recalled by Abbott Laboratories, Inc Due to The Peristaltic Head tubing placed on the ARCHITECT...

Date: September 29, 2017
Company: Abbott Laboratories, Inc
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Laboratories, Inc directly.

Affected Products

ARCHITECT c4000 part number 02P24; ARCHITECT c8000 part number 01G06; ARCHITECT c16000 part number 03L77 Clinical Chemistry systems.

Quantity: 34 units

Why Was This Recalled?

The Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector.

Where Was This Sold?

Worldwide distribution to US, Canada, France, and Germany.