Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

"HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 Recalled by Abbott Laboratories Due to Calibrator level 2 bottle may have been manufactured...

Date: May 22, 2019
Company: Abbott Laboratories
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Laboratories directly.

Affected Products

"HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 *** FOR USE WITH ARCHITECT" - Product Usage: use in the calibration of the Hemoglobin A1c assay.

Quantity: 198 Kits

Why Was This Recalled?

Calibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. The manufacturer observed no impact to assay performance between the HbA1c concentration range of 5.33% to 6.87%. However, outside of this range, a bias of greater than 3% may be seen with patient results. The device failure that is associated with this recall can result in potentially negatively and positively biased patient test results for HbA1c. The firm is investigating the root cause.

Where Was This Sold?

This product was distributed nationwide across all 50 states.