Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Emerald Diluent Tubing Recalled by Abbott Laboratories Due to The diluent tubing used on the instrument appears...

Date: May 9, 2019
Company: Abbott Laboratories
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Laboratories directly.

Affected Products

Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.

Quantity: 22 analyzers with tubing

Why Was This Recalled?

The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.

Where Was This Sold?

This product was distributed to 7 states: GA, LA, MA, NY, NC, SC, TX