Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: Recalled by Abbott Laboratories Due to When expired reagents are scanned or manually entered,...

Date: March 18, 2025
Company: Abbott Laboratories
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Laboratories directly.

Affected Products

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby

Quantity: 45

Why Was This Recalled?

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Where Was This Sold?

This product was distributed nationwide across all 50 states.