Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Alinity hq Analyzer Recalled by Abbott Laboratories Due to 1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic)...

Date: June 14, 2024
Company: Abbott Laboratories
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Laboratories directly.

Affected Products

Alinity hq Analyzer, REF: 09P68-01

Quantity: 816

Why Was This Recalled?

1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts

Where Was This Sold?

This product was distributed nationwide across all 50 states.