Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer Recalled by Abbott Laboratories Due to Due to potential performance issues with software version...

Date: July 26, 2021
Company: Abbott Laboratories
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Laboratories directly.

Affected Products

Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer for in vitro diagnostic use. Model number LN 03R70-01

Quantity: 4,833 systems

Why Was This Recalled?

Due to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There is the potential for onboard reagents past their lot expiration date or onboard stability time to be incorrectly displayed on the Reagent Status screen with a status of "OK". In this scenario, the expired reagent will remain in the carousel for sample processing. 2) Performing monthly maintenance procedure 5701 Clean ICT Drain Tip incorrectly may potentially cause issues such as damaged connector or leaking connection. 3) A sample exception with message code 150 "Unable to process test. Previous processing module error" is produced without notifying the operator of a reagent pipettor failure and may cause processing module to transition into the Pausing state. Tests initiated prior to the sample exception may continue to process without dispense of reagents, potentially leading to incorrect results. 4. Reagent short dispense: Alinity c needs a control measure to protect against the potential for short reagent dispense. Addition of a maximum depth limit for reagent cartridges on the Alinity c to prevent the potential for a short dispense is needed. 5. RSM indicators were blinking green incorrectly: The Reagent and Sample Manager (RSM) indicators will incorrectly blink green indicating that processing is completed and the rack or cartridge can be removed from the loading area when the rack or cartridge should not be accessed. 6. Erroneous ICT Calibration Curves: Erroneous ICT calibration curves might be generated for the Alinity c when there is a malfunction of ICT reference solution delivery, use of wrong/evaporated calibrator, defective ICT modules, etc.

Where Was This Sold?

This product was distributed nationwide across all 50 states.