Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARCHITECT Total T3 Reagent Kit (7K64) consists of: - 1 Recalled by Abbott Ireland Diagnostics Division Due to 17% of ARCHITECT Total T3 Reagent Kit, LN...

Date: September 11, 2014
Company: Abbott Ireland Diagnostics Division
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Ireland Diagnostics Division directly.

Affected Products

ARCHITECT Total T3 Reagent Kit (7K64) consists of: - 1 or 4 Bottle(s) (6.6 mL/27.0 mL) anti-T3 (sheep) coated microparticles in MES buffer with sheep IgG stabilizers. Minimum Concentration: 0.08% solids. Preservative: ProClin 300. - 1 or 4 Bottle(s) (5.9 mL/26.3 mL) T3 acridinium-labeled conjugate in citrate buffer with NaCl and Triton X-100 stabilizers. Minimum concentration: 0.33 ng/mL. Preservative: ProClin 300. The ARCHITECT Total T3 (TT3) assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of total triiodothyronine (Total T33) in human serum and plasma.

Quantity: 7,415 kits

Why Was This Recalled?

17% of ARCHITECT Total T3 Reagent Kit, LN 7K64-20/-25 reagent lots 38901UI00 and 38901UI01 may exhibit lower Relative Light Units (RLUs) than expected, which in turn may result in controls out of range or patient results higher than expected.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Abbott Ireland Diagnostics Division

Abbott Ireland Diagnostics Division has 12 total recalls tracked by RecallDetector.