Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained Recalled by Abbott Gmbh & Co. KG Due to Abbott identified that samples tested for ARCHITECT STAT...

Date: April 24, 2020
Company: Abbott Gmbh & Co. KG
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Gmbh & Co. KG directly.

Affected Products

ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.

Quantity: 6087 Reagent Kits

Why Was This Recalled?

Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Abbott Gmbh & Co. KG

Abbott Gmbh & Co. KG has 18 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report