Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Abbott Alinity c Processing Module LN 03R67-01 - Product Usage: Recalled by Abbott Gmbh & Co. KG Due to The safety interlock covering the septum piercing probes...

Date: February 26, 2019
Company: Abbott Gmbh & Co. KG
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Gmbh & Co. KG directly.

Affected Products

Abbott Alinity c Processing Module LN 03R67-01 - Product Usage: The Alinity c processing module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. The Alinity c processing module uses photometric detection technology to measure sample absorbance for the quantification of analyte concentration and uses potentiometric detection technology to measure the electrical potential in a sample. In addition, the Alinity c processing module uses an integrated chip technology (ICT) module to measure potentiometric assays (electrolytes).

Quantity: 559

Why Was This Recalled?

The safety interlock covering the septum piercing probes within the bulk solution bottle holder may not deploy when a bulk solution bottle is removed from the Alinity ci series and Alinity s System.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Abbott Gmbh & Co. KG

Abbott Gmbh & Co. KG has 18 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report