Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Power supply provided with Afinion 2 Recalled by Abbott Diagnostics Technologies AS Due to The analyzer system may have an electrostatic discharge...

Date: March 23, 2020
Company: Abbott Diagnostics Technologies AS
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Diagnostics Technologies AS directly.

Affected Products

Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970, 1116971, 1116985, 1116986, IVD. The firm name on the label is Abbott Diagnostics Technologies AS, Oslo, Norway. Power supply provided with Alere Afinion AS100 Analyzer, Catalog numbers 1115175, 1115390, 1116049, 1116050, 1116053, 1116054, 1116456, and 1116980. The firm name on the label is Alere Technologies AS, Oslo, Norway.

Quantity: 42,237 instruments

Why Was This Recalled?

The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the analyzers resulting in the power supply becoming non-functional.

Where Was This Sold?

This product was distributed to 1 state: NJ

Affected (1 state)Not affected

About Abbott Diagnostics Technologies AS

Abbott Diagnostics Technologies AS has 2 total recalls tracked by RecallDetector.