Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

FreeStyle Flash Blood Glucose Monitors System. Meter made in China Recalled by Abbott Diabetes Care, Inc. Due to Abbott Diabetes Care has identified through internal testing...

Date: February 19, 2014
Company: Abbott Diabetes Care, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Abbott Diabetes Care, Inc. directly.

Affected Products

FreeStyle Flash Blood Glucose Monitors System. Meter made in China; Distributed by TheraSense, Alameda, CA The FreeStyle Flash meter should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Meter can product inaccurate results. Intended to monitor blood glucose from samples taken from the body. IVD use only,

Quantity: 25,203 units

Why Was This Recalled?

Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.

Where Was This Sold?

This product was distributed nationwide across all 50 states.