Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7 Recalled by 89 NORTH Due to The defect relates to firmware version lower than...

Date: January 31, 2020
Company: 89 NORTH
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact 89 NORTH directly.

Affected Products

LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging

Quantity: 188

Why Was This Recalled?

The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment only which overrides all interlocks. When this mode - "Interlock Override Mode" - is entered, the system does not cease laser emission when interlocks are opened, but rather allows emission to continue with open interlocks. The potential for this to occur is non-compliant with 21CFR 1040 .10{f )(2) and 21 CFR 1040 .10(f)(3).

Where Was This Sold?

Worldwide

About 89 NORTH

89 NORTH has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report