Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

3M Bair Hugger(TM) Normothermia System Recalled by 3M Company - Health Care Business Due to During a recent investigation, 3M confirmed that a...

Date: October 17, 2017
Company: 3M Company - Health Care Business
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact 3M Company - Health Care Business directly.

Affected Products

3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)

Quantity: 153 cases (3825 sensors)

Why Was This Recalled?

During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.

Where Was This Sold?

This product was distributed to 6 states: CA, GA, IL, MI, MN, MO